The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
Source facts
Action
Siemens sent a Customer Safety Advisory Letter to all affected customers. The letter identified the affected product, problem and actions to be taken. Siemens service organization will be contacting customers to schedule the update as per the attached Update Instruction and will be provided the replacement bracket to resolve the issue. For questions contact your Siemens representative.
Initiated
2016-04-25
Root cause
Process design
Terminated
2017-02-01
Distribution
Worldwide Distribution - US Nationwide and the countries of China and Malaysia.
Recall number
Z-1999-2016
Recall status
Terminated
CDRH record ID
146915
FDA classified
2016-06-15
Recall event number
74239