Z-2001-2026: i-STAT G3+ cartridge; List Number: 03P78-26;

Recall record Z-2001-2026 for Abbott Point Of Care Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2001-2026
Company in FDA record
Abbott Point Of Care Inc.
FDA classification date
Apr 29, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

Source facts

Action
On April 1, 2026, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. This Urgent Medical Device Corrective is to customers in the U.S. with the following recommended actions: " Discontinue use of the affected i-STAT EG7+, EG6+, and G3+ cartridge lots for reporting PCO and pH results. " Use an alternate cartridge type, alternate lot, or testing method. " Complete and return the Business Reply Card included with this notification to confirm receipt and understanding. If you have forwarded any of the specified i-STAT EG7+, EG6+ or G3+ cartridges to another facility, we request that you please provide a copy of this letter to them.
Initiated
2026-04-01
Root cause
Process change control
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.
Recall number
Z-2001-2026
Recall status
Open, Classified
CDRH record ID
219278
FDA classified
2026-04-29
Recall event number
98635

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗