Open original FDA record ↗Recall record
- FDA identifier
- Z-2002-2012
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Jul 12, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports.
Source facts
- Action
- IMPAC Medical Systems, Inc. sent an Important Safety Notice LINFFM0001/1.0 on June 21, 2012, to consignees on concerning a patch to the software. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to not use the DICOM coordinates option for Reports when printing the Setup Reference Diaglog. Customers were instructed go to www.elekta.com, select the SupportPlus Login at the tope of the page, enter their portal, select Downloads/Updates to download the latest patches. For questions regarding this recall call 770-670-2548.
- Initiated
- 2012-06-21
- Root cause
- Software design
- Terminated
- 2014-04-01
- Distribution
- Nationwide Distribution including CA, CT, FL, IL, IN, KY, MI, MN, MO, NJ, NY, NC, OH, OK, OR, PA, WA, and WI.
- Recall number
- Z-2002-2012
- Recall status
- Terminated
- CDRH record ID
- 110552
- FDA classified
- 2012-07-12
- Recall event number
- 62433