Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.
Source facts
Action
The firm began notifying its sales representatives on 06/02/2021 by email or hand delivery and followed with notifying the implanting/following physicians and centers on 06/03/2021 by letter. The letter explained the issue and provided recommendations for the physicians. Information is also posted on www.BostonScientific.com/advisory. Patient notification is done by the implanting/following physician.
Initiated
2021-06-03
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution.
Recall number
Z-2002-2021
Recall status
Open, Classified
CDRH record ID
187854
FDA classified
2021-07-02
Recall event number
88087