Z-2003-2020: Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - P

Recall record Z-2003-2020 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2003-2020
Company in FDA record
Inpeco S.A.
FDA record creation date
May 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.

Source facts

Action
On April 7, 2020, the firm sent an Urgent Medical Device Correction letter to all impacted customers and instructed them to return the Customer Letter Receipt Confirmation and Implementation Check form within 15 days. The customers were asked to confirm that they had read, understood, and applied all the applicable recommended actions. The impacted Customers were told that they would be contacted by their service representative to schedule the system update.
Initiated
2020-04-07
Root cause
Software design
Terminated
2022-07-28
Distribution
Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington
Recall number
Z-2003-2020
Recall status
Terminated
CDRH record ID
180944
FDA record created
2020-05-14
Recall event number
85454

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗