- Recall classification
- Class II
Reason stated in the source
When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
Source facts
- Action
- On April 7, 2020, the firm sent an Urgent Medical Device Correction letter to all impacted customers and instructed them to return the Customer Letter Receipt Confirmation and Implementation Check form within 15 days. The customers were asked to confirm that they had read, understood, and applied all the applicable recommended actions. The impacted Customers were told that they would be contacted by their service representative to schedule the system update.
- Initiated
- 2020-04-07
- Root cause
- Software design
- Terminated
- 2022-07-28
- Distribution
- Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington
- Recall number
- Z-2004-2020
- Recall status
- Terminated
- CDRH record ID
- 180945
- FDA record created
- 2020-05-14
- Recall event number