When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.
Source facts
Action
Permobil began notifying customers via in-person and through letters on 05/10/2023. Customers were instructed to examine their device and update their software to SmartDrive MX2+ Version 1.1.00, to acknowledge the update was conducted and the notification letter was understood through the field action portal, notify customers if the affected devices were further distributed. They were also instructed to discontinue using the affected device while waiting for the correction.
Initiated
2023-05-10
Root cause
Software design
Distribution
Worldwide distribution.
Recall number
Z-2004-2023
Recall status
Open, Classified
CDRH record ID
199997
FDA classified
2023-06-26
Recall event number
92250