Skip to content

Public FDA record

Z-2005-2021: Recirculator 8.0s Disposable Lavage Kit (product code number 8100) , GTIN: 08719992086016. Hyperthermic

Recall record Z-2005-2021 for Eight Medical International BV. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2005-2021
Company in FDA record
Eight Medical International BV
FDA record creation date
Jul 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The patient may be exposed to potential aluminum ion release during warming therapy under certain use conditions.

Source facts

Action
An Urgent: Medical Device Recall notification letter dated 6/7/21 was emailed and mailed to customers. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: Inspect inventory and quarantine impacted product. Remove and destroy the instruction for use (IFU) (Rev F) upon opening the kit during the treatment Replace with the IFU (Rev G) attached to this letter Attach the laminated card provided with this letter on the fluid warmer device itself, not on the disposable kit. Until we are able to provide kits with the corrected IFU (Rev G), please continue to use the attached Rev G. Complete and return response form FM-018-02 via email to hurtt.tracy@eightmedical.com (the document is attached) TYPE OF ACTION BY THE COMPANY: EIGHT MEDICAL INTERNATIONAL B.V. is working towards an updated design solution and will notify customers in order to replace products, when available. OTHER INFORMATION: If you need any further information or support concerning this issue, please contact Tracy Hurtt, hurtt.tracy@eightmedical.com For calls +1 (812) 212-2198 - Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time. For Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2021-06-04
Root cause
Device Design
Terminated
2023-08-03
Distribution
US Nationwide, Colombia, and Dubai.
Recall number
Z-2005-2021
Recall status
Terminated
CDRH record ID
185975
FDA record created
2021-07-23
Recall event number
87406

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗