Sufficiently worn/damaged Universal Cable with product number MC5057, when used with various ConMed Linvatec electric Handpieces, may cause the handpiece to self activate.
Source facts
Action
ConMed Linvatec notified customers via letter, cautioning the users to inspect cords for signs of excessive wear or damage and urging them to discontinue use and return cables when appropriate. Direct questions to Linvatec Corporation by calling 1-800-237-0169.
Initiated
2009-07-31
Root cause
Component design/selection
Terminated
2011-07-11
Distribution
Worldwide Distribution including United States.
Recall number
Z-2007-2009
Recall status
Terminated
CDRH record ID
84057
FDA classified
2009-10-01
Recall event number
52812
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.