Z-2007-2026: Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case

Recall record Z-2007-2026 for Medical Action Industries, Inc. 306. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2007-2026
Company in FDA record
Medical Action Industries, Inc. 306
FDA classification date
May 6, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"

Source facts

Action
On April 22, 2026, Medial Action Industries, Inc. (subsidiary of Owens & Minor) issued a "Urgent: Medical Device Recall Notification Update" via E-Mail to affected consignees. The updated notification provided clarity and and additional information for the recall strategy. Initial notification was issued on March 17, 2026. Medical Action Industries asked consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product. Convenience kits can still be used once properly labeled. 2. Add warning labels to affected Medical Action kits to ensure all users are aware of the actions below and identified in the Medline recall notice. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. 3. If your firm used the labeling template provided with the initial letter, please adhere the revised label over the existing template label. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by April 30, 2026. If you have any questions, contact Reg. Affairs Manager, Kitting, Global Products, by email: GM-OMRA_RECALLS@owens-minor.com or 470-280-4277.
Initiated
2026-03-17
Root cause
Unknown/Undetermined by firm
Distribution
US distribution to state of: AZ OUS: None
Recall number
Z-2007-2026
Recall status
Open, Classified
CDRH record ID
219499
FDA classified
2026-05-06
Recall event number
98698

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗