Z-2008-2008: DST-XL Model 1002375 is a multi-head, whole body gamma camera with infrared body contouring and anti-coll

Recall record Z-2008-2008 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2008-2008
Company in FDA record
GE Healthcare
FDA classification date
Sep 5, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

GE Healthcare has recently became aware of a cable failure associated with the arm or gantry rotation of the DST-XL Camera that may impact patient safety.

Source facts

Action
Field Modification Instructions (FMI No: 40815) will be issued to the field for the DST-XL product to install an arm or gantry rotation motor's brake. The customers letters "Product Safety Notification" dated March 10, 2008, have been distributed to Hospital Administrators, Mangers of Radiology/Cardiology and Radiologists/Cardiologists. The letter informs the customers of the safety issue, affected product, safety instructions, product correction and contact information. If you have questions, please contact GE Healthcare within the US or Canada at 1-800-437-1171 or outside the US the telephone number at 1-262-896-2890.
Initiated
2008-03-10
Root cause
Device Design
Terminated
2011-12-10
Distribution
AL, AR, CA , CO, DE , FL, IA, IL, KS, MA, MI, MS, NC , NJ, NY, OH, OK, PA , SC, TN, TX , VA, VT, WA United Kingdom, Belgium, Switzerland, Russia Federal Sweden, France, Italy, Spain, Germany Federal, Norway, Portugal, Netherlands, Saudi Arabia, Iran, India, Canada, Argentina, Taiwan, China, Bangladesh, Austria.
Recall number
Z-2008-2008
Recall status
Terminated
CDRH record ID
70893
FDA classified
2008-09-05
Recall event number
48137

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗