Skip to content

Public FDA record

Z-2008-2009: Model 100 enFlow Fluid/Blood Warmer Model Numbers: 980100, 91000103, 91000153, 91000154, 91000156 The

Recall record Z-2008-2009 for Enginivity LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2008-2009
Company in FDA record
Enginivity LLC
FDA classification date
Sep 4, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

1. Fluid contamination of the electronic circuitry can cause short circuits resulting in permanent fuse failure or localized circuit board overheating with possible smoldering and discoloration. 2. Electromagnetic (EMI) Interference associated with the Warmer may produce artifact in ECG, EEG, or EMG records (cardiac or neuro monitoring).

Source facts

Action
Vital Signs, Inc. issued Product Safety Notification on March 23, 2009 informing users of potential EMI and fluid ingress issues. All warmers in the field will be replaced beginning in May 2009.
Initiated
2009-03-23
Root cause
Device Design
Terminated
2012-05-02
Distribution
Worldwide Distribution -- USA, Philippines, Saudi Arabia, India, Israel, Australia, Mexico, Germany, Republic of Singapore, China, Guatemala, Switzerland, Dubai United Arab Emirates, New Zealand, Malaysia, India, Ecuador, & United Kingdom.
Recall number
Z-2008-2009
Recall status
Terminated
CDRH record ID
82524
FDA classified
2009-09-04
Recall event number
52209

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.