Open original FDA record ↗Recall record
- FDA identifier
- Z-2008-2009
- Company in FDA record
- Enginivity LLC
- FDA classification date
- Sep 4, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
1. Fluid contamination of the electronic circuitry can cause short circuits resulting in permanent fuse failure or localized circuit board overheating with possible smoldering and discoloration. 2. Electromagnetic (EMI) Interference associated with the Warmer may produce artifact in ECG, EEG, or EMG records (cardiac or neuro monitoring).
Source facts
- Action
- Vital Signs, Inc. issued Product Safety Notification on March 23, 2009 informing users of potential EMI and fluid ingress issues. All warmers in the field will be replaced beginning in May 2009.
- Initiated
- 2009-03-23
- Root cause
- Device Design
- Terminated
- 2012-05-02
- Distribution
- Worldwide Distribution -- USA, Philippines, Saudi Arabia, India, Israel, Australia, Mexico, Germany, Republic of Singapore, China, Guatemala, Switzerland, Dubai United Arab Emirates, New Zealand, Malaysia, India, Ecuador, & United Kingdom.
- Recall number
- Z-2008-2009
- Recall status
- Terminated
- CDRH record ID
- 82524
- FDA classified
- 2009-09-04
- Recall event number
- 52209