Open original FDA record ↗Recall record
- FDA identifier
- Z-2008-2012
- Company in FDA record
- DeRoyal Industries Inc
- FDA classification date
- Jul 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Surgical kits and trays contain Stryker Hytrel Togas which were recalled because the clear tape, that is intended to aid in the prevention of patient's fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process.
Source facts
- Action
- DeRoyal initiated the recall on 6/16/2012 by notifying customers by overnight delivery of letters describing the problem, instructing the customer how to handle inventory on hand, and requesting the customer to destroy the product, fill out the provided notice of destruction form, and return the form to DeRoyal. For questions customers were instructed to call 865-362-1020 or 865-362-1034. For questions regarding this recall call 865-362-2334.
- Initiated
- 2012-06-16
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-11-09
- Distribution
- Nationwide Distribution including SD and NE.
- Recall number
- Z-2008-2012
- Recall status
- Terminated
- CDRH record ID
- 110698
- FDA classified
- 2012-07-13
- Recall event number
- 62457