Z-2010-2008: Physio-Control LIFEPAK 20e defibrillator/monitor. This product is an external defibrillator sold interna

Recall record Z-2010-2008 for Physio Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2010-2008
Company in FDA record
Physio Control, Inc.
FDA classification date
Sep 8, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A thicker keypad that may prevent the door from fully latching closed. If the door is not fully closed, there is a potential that the defibrillator will not automatically change from automatic external defibrillation (AED) mode to manual mode by pressing the "MANUAL" button located on the lower left corner of the door. This may lead to a delay in defibrillation therapy.

Source facts

Action
On 5/13/08, the firm sent out letters to their domestic customers via Certified Receipt. The letter advised customers of the recall and instructed them to do the following, "Keep the defibrillator in service. If pressing the MANUAL button does not take the defibrillator out of AED mode, press one of the following keys to select manual mode: "ENERGY SELECT", "CHARGE", "PACER", OR "LEAD"... Please call your local Physio-Control service representative to arrange a service visit to replace the keypads." If you have questions, contact Technical Support at 1-800-442-1142, option 5.
Initiated
2008-05-13
Root cause
Device Design
Terminated
2011-03-20
Distribution
Product distributed to 1 US embassy, 97 domestic medical facilities nationwide, and 5 foreign consignees including the following countries: Germany, Netherlands, Canada, Australia and Hong Kong.
Recall number
Z-2010-2008
Recall status
Terminated
CDRH record ID
71374
FDA classified
2008-09-08
Recall event number
48139

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗