Z-2010-2021: Verigene Clostridium difficile Nucleic Acid Test Kit (CDF) Model Number: 20-006-022

Recall record Z-2010-2021 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2010-2021
Company in FDA record
Luminex Corporation
FDA record creation date
Jul 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for a false-negative result.

Source facts

Action
On October 2020 the firm sent a letter to all their consignees with the following information: Dear Valued Luminex Customer, According to our records, you have ordered and may have used our portfolio of VERIGENE Assays in your laboratory. Through post-market surveillance activities, we have identified an increase in the No Call/False Positivity rate for these products. Our investigation has identified that the likelihood of a No Call/False Positive for the assays below increases when test cartridges are left on the Processor SP for an extended time following completion of assay preparation. We are currently working on a software solution aimed at further optimizing the washing and drying steps in the sample preparation process. This software update should minimize No Call/False Positive rates for these assays. Until this solution is finalized and implemented, we advise you to remove test cartridges from the VERIGENE SP Processor immediately following completion of assay preparation, and then to process these cartridges on the VERIGENE Reader as soon as possible to obtain test results. While this is common practice for many customers already, doing so should minimize the occurrence of No Call/False Positive results for the products listed below. We will contact you with more details on the software solution soon as possible. Please contact Luminex Global Support Services with any questions or concerns. Luminex Global Support Services 1-877-785-2323 (U.S. and Canada) +1-512-381-4397 (Outside U.S. and Canada) support@luminexcorp.com
Initiated
2020-10-30
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide including in the states of AL, AZ, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA and VT. The countries of Austria, Chile, France, Germany, Greece, Italy, Japan, Kuwait, Netherlands, Spain, Thailand, Turkey, and United Kingdom.
Recall number
Z-2010-2021
Recall status
Open, Classified
CDRH record ID
188010
FDA record created
2021-07-02
Recall event number
88137

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗