Z-2012-2021: ACE Control Set - IVD Control set for the quantitative determination of Angiotensin Converting Enzyme (AC

Recall record Z-2012-2021 for Sentinel CH SpA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2012-2021
Company in FDA record
Sentinel CH SpA
FDA record creation date
Jul 6, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Claim for stability after reconstitution from lyophilized to liquid form does not meet the requirements. New IFU instructions for handling the product provided to end users.

Source facts

Action
Sentinel notified distributors by email on the 30th April 2021 and provided the letter (FSN) to be forwarded to final end-users. Letter states reason for recall, health risk and action to take: Actions to be taken: 1. In order to use the product properly, please follow the instructions described here below. 2. Fill in all the required fields of the "Receipt of the Safety Notice" and return it to Sentinel Diagnostics; 3. Review the content of this communication with your Medical Director and retain this letter for any future reference. New procedure Take the Control REF 1667001 from the refrigerator. Carefully remove the cap. " Gently remove the stopper to avoid loss of the lyophilized pellet and add exactly 1.0 ml of distilled or deionized water. Replace stopper and gently swirl. " Let to stand for 24 hrs at 2-8C. Swirl gently each vial immediately before use, avoiding foaming. " Take the volume necessary for the scheduled analysis. Recap and store at 2-8 C. Stability after reconstitution: 5 days when stored at 2- 8 C if contamination is avoided and vials are recapped immediately after use. Transmission of this Field Safety Notice: Please forward this notice to all the indivuals in your organization who need to be aware or to any organization/individuals where the potentially affected devices have been transferred. Reference person: If you or any of your customers have any questions regarding this information, please contact your local area Customer Service
Initiated
2021-04-30
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, CANADA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, JORDAN, NEW ZEALAND, POLAND, PORTUGAL, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, THE NETHERLANDS, TURKEY, U.A.E., UK.
Recall number
Z-2012-2021
Recall status
Open, Classified
CDRH record ID
187740
FDA record created
2021-07-06
Recall event number
88040

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗