Skip to content

Public FDA record

Z-2013-2021: Randox RX Imola Analyser with ISE, Catalog Numbers RX4900, RX4900R

Recall record Z-2013-2021 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2013-2021
Company in FDA record
Randox Laboratories Ltd.
FDA record creation date
Jul 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue was identified where the software froze during processing of commands, which resulted in no results displayed.

Source facts

Action
Urgent Medical Device Correction letters dated 4/28/21 were sent to customers. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. " Please contact your local service provider to arrange for the software update to be performed . "Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. Please accept our apologies for any inconvenience caused. Thank you for your patience and understanding. If you have any questions or concerns please contact Randox Technical Services. Local Contact Information USA Randox Laboratories-US, Ltd. 515 Industrial Boulevard Kearneysville West Virginia, 25430 Tel: +1 304 728 2890 Toll Free 866 4 RANDOX Email: customersupportusa@randox.com Puerto Rico Clinical Diagnostics of P.R. LLC Carr. 887km 0.6, Carolina Commercial Park Local B-2 Carolina, Puerto Rico Tel: 1-787-701-7000 Email: Jose.Rosario@randox.com
Initiated
2021-04-28
Root cause
Software design
Terminated
2021-12-03
Distribution
US Nationwide Distribution and Puerto Rico.
Recall number
Z-2013-2021
Recall status
Terminated
CDRH record ID
187859
FDA record created
2021-07-26
Recall event number
88098

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗