Open original FDA record ↗Recall record
- FDA identifier
- Z-2014-2012
- Company in FDA record
- Etac Supply Center Ab
- FDA classification date
- Jul 14, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The recall has been initiated because there is the potential that users may fall and become injured, with the potential for serious injury, including bone fracture. One death occurred in a user fall incident, but it is unclear whether the fall was caused by the loss of a wheel. Among the walkers distributed worldwide, Etac has received a total of 49 reports from Europe and Japan of wheels falling
Source facts
- Action
- Etac sent an "URGENT DEVICE RECALL " letter dated March 20, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to immediately discontinue selling and distributing the affected product. A Reply Postcard was enclosed for customers to complete and return. Contact the firm at 847-277-7705 for questions regarding this recall.
- Initiated
- 2012-04-06
- Root cause
- Device Design
- Terminated
- 2013-06-18
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, China, Colombia, Germany, Denmark, Estonia, Spain, Finland, France, Great Brittain, Greenland, Greece, Hungaria, Ireland, Israel, Iceland, Italy, Japan, South Korea, Kuwait, Lithuania, Luxembourg, Latvia, Malaysia, Netherlands, Norway, Poland, Portugal, Rumania, Russia, Saudi Arabia, Sweden, Slovenia, Slovakia, and Thailand.
- Recall number
- Z-2014-2012
- Recall status
- Terminated
- CDRH record ID
- 108163
- FDA classified
- 2012-07-14
- Recall event number
- 61463