Z-2017-2025: Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyx

Recall record Z-2017-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2017-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Jun 24, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.

Source facts

Action
On 5/8/2025, correction notices were mailed and emailed to Customers informing them of the following: Firm has temporarily configured impacted devices to "disabled" mode so the device will not automatically apply validated patches intended to address security vulnerabilities or performance issues. Device keys can be used to manually access unavailable drawers and must be readily available and retrievable to constitute risk mitigation; medications can also be obtained from other unaffected dispensing devices or pharmacy. The Pyxis Anesthesia Station unlocks non-secure drawers upon user login, providing immediate access to medications. Non-secure drawers remain unlocked while the user is logged in. User guides include instructions on how to manually access medications in an emergency: BD Pyxis MedStation ES user guide DME 10000430667 version 00 (page 120), BD Pyxis Anesthesia Station ES user guide DME 10000430730 version 00 (page 84). Other potential mitigations: -Healthcare facility system administrators can grant permission to add temporary patients to select staff per user guide DME 10000430667 version 00 (page 25). -Critical override or temporary Non-RX mode can be configured for impacted profile devices per user guide DME 10000430667 version 00 (pages 74-75). -Refill reports can be generated at the device to identify items requiring replenishment if inventory information is not communicated to the server. Share the recall letter with all network facilities and post recall letter on/near affected products. Complete and return response form: BDRC22@bd.com Firm will coordinate with customers to obtain a schedule to implement "Managed Mode". Firm contacts: North American Regional Complaint Center: 1-844-823-5433, productcomplaints@bd.com; BD Remediation Support: 1-866-583-8783; BD Technical Support: 1-800-727-6102 On 12/2/2025 updated correction notices including additional affected devices were mailed/emailed to customers
Initiated
2025-05-08
Root cause
Software Design Change
Distribution
Worldwide - US Nationwide distribution including in the states of PR, DE, GU, MO, IA, IL, CA, MD, MA, VA, PA, NY, WA, NJ, AZ, CO, OH, FL, WI, TX, CT, LA, MI, WY, MN, AR, KY, OR, NM, TN, NC, SC, IN, VT, KS, AL, GA, WV, NV, ID, DC, MS, NE, HI, OK, UT, ND, MT, NH, SD, ME, RI, AK, MP and the countries of SA, CA, NZ , CH, FR, IT, BE, AU, BH, QA, AE, GB, EG, AR, JO, DE, BM, TH, LB, BR, IE, FI, MX, TW, ID, SG, BS, IN, MA, GF, ES, HK.
Recall number
Z-2017-2025
Recall status
Open, Classified
CDRH record ID
213211
FDA classified
2025-06-24
Recall event number
96585

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗