Skip to content

Public FDA record

Z-2018-2021: INVIVE CRT-P, Model Numbers: a) V172 b) V173 c) V182 d) V183 e) W172 f) W173

Recall record Z-2018-2021 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2018-2021
Company in FDA record
Boston Scientific Corporation
FDA classification date
Jul 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population

Source facts

Action
The firm issued an advisory communication on June 3, 2021 by letter to the implanting/following physicians and clinics regarding the potential device behavior. The notice included a letter to the patients to be disseminated to the patients implanted with the device at the discretion of the physician. BSC also made the patient letter available on their advisory page, found on our product performance website @ https://www.bostonscientific.com/en-US/pprc/product-advisories.html
Initiated
2021-06-03
Root cause
Device Design
Distribution
Worldwide distribution.
Recall number
Z-2018-2021
Recall status
Open, Classified
CDRH record ID
187902
FDA classified
2021-07-13
Recall event number
88107

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗