Z-2019-2024: VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in veri

Recall record Z-2019-2024 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2019-2024
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Jun 4, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.

Source facts

Action
QuidelOrtho issued IMPORTANT PRODUCT CORRECTION NOTIFICATION (Ref. CL2024-123) to Distributors and End-Users via FedEx overnight courier and/ or electronically. Letter states reason for recall, health risk and action to take: " Discontinue using, render unusable, and discard your remaining inventory of VITROS Vitamin B12 Range Verified, Lot 1200. QuidelOrtho will replace or credit your account. Indicate quantities to be replaced or credited via the Confirmation of Receipt form. " Complete the enclosed Confirmation of Receipt form no later than May 14, 2024. " Please forward this notification if the affected product was distributed outside of your facility. If you have further questions, please contact the Global Services Organization at 1-800-421-3311.
Initiated
2024-05-08
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide and the country of Columbia.
Recall number
Z-2019-2024
Recall status
Open, Classified
CDRH record ID
207931
FDA classified
2024-06-04
Recall event number
94650

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗