Z-2020-2024: Ventilator HAMILTON-C6, PN: 160021

Recall record Z-2020-2024 for Hamilton Medical AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2020-2024
Company in FDA record
Hamilton Medical AG
FDA classification date
Jun 18, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.

Source facts

Action
On 5/24/24, correction notices were emailed to customers and distributors and they were asked to do the following: If ventilation is not re-initiated due to this issue, there are 4 ways to reinitiate ventilation: 1) Select and confirm a control setting (even without a change of the value), or 2) Select and confirm a new ventilation mode, or 3)Switch the ventilator to standby mode and restart ventilation from standby mode, or 4) Switch the ventilator off and switch it on again If the user is not able to manually re-initiate ventilation in due time as per one of the 4 ways described above, they might decide to ventilate the patient by alternative means. - Read and sign the confirmation sheet, which can be returned to bret.everett@hamiltonmedical.com - Ensure that your medical staff is aware of the notice content - Attach this letter to each affected device's operator's manual in order to ensure quick access. - The newest available software version 1.2.3 contains the required fix and shall be installed without undue delay. Your local distribution partner will support you with updating your affected devices with high priority. Questions, contact techsupport@hamilton-medical.com
Initiated
2024-05-15
Root cause
Packaging change control
Distribution
US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.
Recall number
Z-2020-2024
Recall status
Open, Classified
CDRH record ID
207956
FDA classified
2024-06-18
Recall event number
94671

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗