Skip to content

Public FDA record

Z-2021-2009: Bard * Button Device Continuous Feeding Tube with 90 degree Adaptor, 18F x 1.7cm, NON STERILE, REF 00028

Recall record Z-2021-2009 for C R Bard Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2021-2009
Company in FDA record
C R Bard Inc
FDA classification date
Sep 17, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Possible restriction/occlusion in adaptors/tube accessories causing difficulty to infuse during routine feeding treatment.

Source facts

Action
Customers were notified by letter and instructed to not use or further distribute any of the affected lot numbers. Customers were further instructed to return the recalled products to Bard Access Systems (BAS) through a return goods authorization number. Questions and requests for assistance should be directed to BAS Customer Service at 800-290-1689.
Initiated
2009-01-20
Root cause
Equipment maintenance
Terminated
2009-12-09
Distribution
Worldwide Distribution -- US, Canada, Austria, Australia, Belgium, Finland, France, Germany, Great Britain, Holland, Italy, Norway, Spain, Sweden and Switzerland, Japan, and Latin America.
Recall number
Z-2021-2009
Recall status
Terminated
CDRH record ID
84196
FDA classified
2009-09-17
Recall event number
52862

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗