Open original FDA record ↗Recall record
- FDA identifier
- Z-2021-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 18, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. The potential exists for a cable connection to become overheated, leading to the emission of smoke and possible material damage.
Source facts
- Action
- Siemens Medical Solutions Inc., issued a Customer Safety Advisory Notice dated November 24, 2010, to affected customers via Update Instructions MR029/10S. Affected customers will be notified via hand delivered letter by Siemens Service Engineers at the time of inspection. A future Update instruction will be released for installation of detectors in appropriate locations on systems to switch off the power electronics in the event of a malfunction in the power supply of the gradient coils. Further questions regarding this recall please call (610) 448-3237.
- Initiated
- 2011-03-03
- Root cause
- Other
- Terminated
- 2014-12-09
- Distribution
- Product was distributed domestically to hospitals and imaging centers in the District of Columbia, Puerto Rico, AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY(every State except VT).
- Recall number
- Z-2021-2011
- Recall status
- Terminated
- CDRH record ID
- 98281
- FDA classified
- 2011-04-18
- Recall event number
- 58080