Z-2021-2023: Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the a

Recall record Z-2021-2023 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2021-2023
Company in FDA record
Randox Laboratories Ltd.
FDA classification date
Jun 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range

Source facts

Action
Randox issued via email to the Distributor Medical Device Correction Letter and forwarded to US and PR customers on 04 May, 2023. Letter states reason for recall, health risk and action to take: Discontinue use of and discard the current IFU and download the updated IFU from randox.com Discuss the contents of this notice with your Medical Director. Complete and return the response for 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization. If you have any questions or concerns please contact Randox Technical Services.
Initiated
2023-05-04
Root cause
Nonconforming Material/Component
Distribution
CA, IL, VA, WA, WI including PR
Recall number
Z-2021-2023
Recall status
Open, Classified
CDRH record ID
200354
FDA classified
2023-06-28
Recall event number
92344

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗