Skip to content

Public FDA record

Z-2021-2024: Brand Name: GlideScope Core 15-inch Monitor Model/Catalog Number: 0570-0404 Product Description: Seriali

Recall record Z-2021-2024 for Verathon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2021-2024
Company in FDA record
Verathon, Inc.
FDA classification date
Jun 4, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.

Source facts

Action
On April 30, 2024, Verathon issued "Urgent Medical Device Correction" notification via FedEx. Verathon asked consignees to take the following actions: 1. Plug your GlideScope Core 15 or Core 15 FHD into the power supply and connect to a hospital-grade power outlet. 2. Use the enclosed USB drive with the corrective software upgrade and related instructions to upgrade your GlideScope Core 15 or Core 15 FHD monitor(s) (approximately 5-7 minutes for the upgrade). 3.Verify the software version on the Administrative Settings tab. OR 4.Contact Verathon Customer Care to contact your Territory Manager to perform the upgrade at your facility. 5.Complete the Correction Response Form in this packet and return it to Verathon via email to CSNotifications@verathon.com. While your fleet is being updated to the latest software, you can continue using the GlideScope Core 15 or Core 15 FHD Monitor. Always ensure that alternative airway management methods and equipment are readily available, per the American Society of Anesthesiologists (ASA) guidelines.
Initiated
2024-04-29
Root cause
Software Manufacturing/Software Deployment
Distribution
US: AK AL AR AZ CA CO CT DC FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK ON OR PA QLD RI SC SD TN TX UT VA VIC WA WI WV WY OUS: Hong Kong, Indonesia, Saudi Arabia
Recall number
Z-2021-2024
Recall status
Open, Classified
CDRH record ID
207644
FDA classified
2024-06-04
Recall event number
94547

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗