Open original FDA record ↗Recall record
- FDA identifier
- Z-2021-2025
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jun 27, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
Source facts
- Action
- Siemens Healthineers sent customers a Customer Advisory Notice dated May 15, 2025 informing customers of a potential issue with ARTIS pheno and ARTIS icono systems with the following model numbers: ARTIS pheno 10849000, ARTIS icono biplane 11327600, ARTIS icono ceiling 11328100, ARTIS icono floor 11327700; due to in very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose and appropriate software update will be implemented, based on the current software version of the affected system.
- Initiated
- 2025-05-15
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- Worldwide.
- Recall number
- Z-2021-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214632
- FDA classified
- 2025-06-27
- Recall event number
- 97135