Open original FDA record ↗Recall record
- FDA identifier
- Z-2022-2024
- Company in FDA record
- Verathon, Inc.
- FDA classification date
- Jun 4, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.
Source facts
- Action
- On April 30, 2024, Verathon issued "Urgent Medical Device Correction" notification via FedEx. Verathon asked consignees to take the following actions: 1. Plug your GlideScope Core 15 or Core 15 FHD into the power supply and connect to a hospital-grade power outlet. 2. Use the enclosed USB drive with the corrective software upgrade and related instructions to upgrade your GlideScope Core 15 or Core 15 FHD monitor(s) (approximately 5-7 minutes for the upgrade). 3.Verify the software version on the Administrative Settings tab. OR 4.Contact Verathon Customer Care to contact your Territory Manager to perform the upgrade at your facility. 5.Complete the Correction Response Form in this packet and return it to Verathon via email to CSNotifications@verathon.com. While your fleet is being updated to the latest software, you can continue using the GlideScope Core 15 or Core 15 FHD Monitor. Always ensure that alternative airway management methods and equipment are readily available, per the American Society of Anesthesiologists (ASA) guidelines.
- Initiated
- 2024-04-29
- Root cause
- Software Manufacturing/Software Deployment
- Distribution
- US: AK AL AR AZ CA CO CT DC FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK ON OR PA QLD RI SC SD TN TX UT VA VIC WA WI WV WY OUS: Hong Kong, Indonesia, Saudi Arabia
- Recall number
- Z-2022-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207645
- FDA classified
- 2024-06-04
- Recall event number
- 94547