Skip to content

Public FDA record

Z-2023-2024: BD SARS-CoV-2 Reagents for BD MAX" Systems. RT-PCR test to detect SARS-CoV-2 in nasal, nasopharyngeal, or

Recall record Z-2023-2024 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2023-2024
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Jun 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Assays may produce false negative results due to decreased activity of the enzyme utilized in manufacturing.

Source facts

Action
Consignees were mailed and/or emailed an URGENT: Medical Device Product Correction notification, dated 5/7/24. The notification instructs consignees to immediately inspect their inventory for affected devices and destroy them per the consignee's destruction process, ensure the notification is read and understood, disseminate the notification to among those in your organization and customers who should be made aware, and return the completed Response Form to BD by email to BDRC22@bd.com or by fax to 312-949-0438. Consignees with any complaints can call 1-844-823-5433 or email productcomplaints@bd.com; consignees with any recall related questions can email BDRC22@bd.com.
Initiated
2024-05-07
Root cause
Component design/selection
Distribution
Domestic: Nationwide Distribution; Foreign: AU, BR, BE, CA, CO, & JP.
Recall number
Z-2023-2024
Recall status
Open, Classified
CDRH record ID
207672
FDA classified
2024-06-05
Recall event number
94599

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗