Skip to content

Public FDA record

Z-2024-2012: Respironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous

Recall record Z-2024-2012 for Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2024-2012
Company in FDA record
Respironics, Inc.
FDA classification date
Jul 23, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class I

Reason stated in the source

The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.

Source facts

Action
Respironics, Inc. began contacting its affected US customers by phone beginning on Friday, April 27, 2012, and requested that they quarantine any affected devices within their possession and that they retrieve any devices with patients and quarantine those as well. Replacement devices were to be provided to all affected customers. For questions customers should call 724-387-7651. For questions regarding this recall call 877-387-3311..
Initiated
2012-04-27
Root cause
Component design/selection
Terminated
2012-11-01
Distribution
Worldwide Distribution - USA including AZ, CA, IL, IN, LA, MI, MN, MO, NJ, NY, NC, PA, SC, TN, TX, and WI and the countries of Canada, Denmark, Estonia, Finland, France, Great Britian, India, Italy, Japan, Korea, Saudi Arabia, and Spain.
Recall number
Z-2024-2012
Recall status
Terminated
CDRH record ID
110401
FDA classified
2012-07-23
Recall event number
62393

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗