It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.
Source facts
Action
Luminex issued an Urgent Field Safety Notice - Medical Device Recall notice on 06/05/2023 by email. The notice explained the problem with the stool prep kit, the risk, and requested isolation/quarantine, cease of use, and disposal of the affected products.
Initiated
2023-06-05
Root cause
Other
Distribution
Worldwide and US Nationwide Distribution
Recall number
Z-2024-2023
Recall status
Open, Classified
CDRH record ID
200936
FDA classified
2023-06-28
Recall event number
92451