Open original FDA record ↗Recall record
- FDA identifier
- Z-2025-2009
- Company in FDA record
- Ev3, Inc.
- FDA classification date
- Sep 8, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
ev3 Inc. is conducting a voluntary recall of specific lots of the Protege GPS Self-Expanding Stent System because labeling did not match the actual stent size contained in the package. The affected product was labeled as a 6mm x 80mm Protege GPS stent but actually contained an 8mm x 40mm stent. This situation can lead to vessel damage or inadequate coverage of the target lesion.
Source facts
- Action
- Consignees were emailed a letter with a subject heading of "Medical Device Recall". The e-mail was dated June 10, 2009. The letter informed the consignee of the problem and product involved. They requested that any affected devices remaining in inventory be identified and returned and to fill out and return the "Device Recall Field Action Reconciliation Form".
- Initiated
- 2009-06-10
- Root cause
- Labeling mix-ups
- Terminated
- 2011-12-11
- Distribution
- International Distribution Only -- Brazil, Uruguay, & Malaysia.
- Recall number
- Z-2025-2009
- Recall status
- Terminated
- CDRH record ID
- 84183
- FDA classified
- 2009-09-08
- Recall event number
- 52642