Open original FDA record ↗Recall record
- FDA identifier
- Z-2025-2014
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Jul 9, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
CareFusion is recalling the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip c
Source facts
- Action
- A recall letter is dated 6/25/14 and will be sent out to customers who purchased the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip chamber. The letter informs the customers of the problems identified, potential risk, and actions to be taken. Customers are instructed to promptly complete and return the enclosed mandatory Customer Response Card to acknowledge the receipt of this communication and to expedite the corrective action process.
- Initiated
- 2014-06-17
- Root cause
- Labeling Change Control
- Terminated
- 2016-09-20
- Distribution
- Distributed in the states of FL, NM, WA, WI, and TX.
- Recall number
- Z-2025-2014
- Recall status
- Terminated
- CDRH record ID
- 128305
- FDA classified
- 2014-07-09
- Recall event number
- 68608