Z-2025-2024: BSM-3000 Series Bedside Monitor REF BSM-3572A The Life Scope BSM-3000 Series Bedside Monitoring System i

Recall record Z-2025-2024 for Nihon Kohden America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2025-2024
Company in FDA record
Nihon Kohden America Inc
FDA classification date
Jun 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to software issue the device may give false "SpO2 Probe Failure" alarm

Source facts

Action
On 4/29-30/2024, the firm sent "URGENT: MEDICAL DEVICE CORRECTION" Letters via FedEx to customer informing them that due to a software issue, the SpO2 probe failure message may occur when the value of the resistor built into the SpO2 relay code is incorrectly determined to be outside the normal range. This issue applies to all Life Scope BSM-3572A bedside monitors with software version 08-94. Customer are instructed to: 1. Please notify all individuals that need to be made aware of this issue/correction, and that this Notification should be posted with or near the affected products. Please notify any facilities if any affected product has been transferred to another facility. 2. Complete the attached form and return to Nihon Kohden. 3. If the SpO2 probe failure message is received and the probe appears undamaged, the bedside monitor may be providing an incorrect alarm. This issue may be temporarily resolved by turning off and restarting the BSM-3572A. 4. If adverse reactions or quality problems are experienced with the use of this product, please report to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. 5. If adverse reactions or quality problems are experienced with the use of this product, or if there are any questions or need for further assistance, please contact Nihon Kohden Technical Support at: 1-800-325-0283 or email mailto:vitalsignsTS@nihonkohden.com.
Initiated
2024-04-29
Root cause
Software design
Distribution
U.S. Nationwide distribution in the state of IA.
Recall number
Z-2025-2024
Recall status
Open, Classified
CDRH record ID
207272
FDA classified
2024-06-06
Recall event number
94472

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗