Open original FDA record ↗Recall record
- FDA identifier
- Z-2026-2012
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA classification date
- Jul 18, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.
Source facts
- Action
- Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE Healthcare sent an"URGENT RECALL NOTICE" dated February 22, 2008 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact the firm at 800-874-7378 for questions concerning this recall.
- Initiated
- 2007-11-19
- Root cause
- Device Design
- Terminated
- 2013-01-28
- Distribution
- Worldwide Distribution.
- Recall number
- Z-2026-2012
- Recall status
- Terminated
- CDRH record ID
- 110421
- FDA classified
- 2012-07-18
- Recall event number
- 62404