Z-2026-2024: O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and ped

Recall record Z-2026-2024 for Medtronic Navigation, Inc.-Littleton. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2026-2024
Company in FDA record
Medtronic Navigation, Inc.-Littleton
FDA classification date
Jun 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.

Source facts

Action
Medtronic issued Urgent Medical Device Correction Letter on 4/26/24 to US end-users hand delivered via Field Service representative. OUS communications were initiated with affected consignees on April 17, 2024. Letter states reason for recall, health risk and action to take: A Medtronic Field Service Engineer will contact your facility to schedule time to service your system within 90 days. Required Actions: 1. If your system is actively in use, you may continue to use it pending service by Medtronic. If your system exhibits the issue, it will automatically reset after beeping. If the frequency of the issue impacts your ability to ensure the system can be in place for procedures, an alternate imaging system should be considered. 2. If your system displays a placard identifying this issue, the placard must remain attached to the system. Your Medtronic Field Service Engineer will remove the placard upon completion of service. 3. Use the enclosed Customer Confirmation Form to confirm receipt of this notification letter. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. 4. This notice needs to be passed on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Additional Information: If you have any questions related to this issue, please contact your Medtronic Field Service Engineer or Technical Services for help at 1-888-826-5603 (24 hours/ 7 days per week) or email at rs.navtechsupport@medtronic.com.
Initiated
2024-04-26
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, MN, PR, TX, WI and the countries of Bangladesh, Canada, Costa Rica, India, Mexico, Panama, Philippines.
Recall number
Z-2026-2024
Recall status
Open, Classified
CDRH record ID
207582
FDA classified
2024-06-06
Recall event number
94566

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗