Skip to content

Public FDA record

Z-2027-2024: GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT,

Recall record Z-2027-2024 for GE Medical Systems Information Technologies Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2027-2024
Company in FDA record
GE Medical Systems Information Technologies Inc
FDA classification date
Jun 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.

Source facts

Action
Recall letters dated and issued on 4/9/2024 by the recalling firm explained the safety issue and actions to be taken by the customer/user. Those actions were to not use the affected products listed in the Affected Product Details section of the letter, use an alternate cuff for NIBP monitoring, and destroy all affected product per the customer facility procedures. A photograph of the product was embedded in the letter to show where to find the product name, model number, and GTIN number. The letter informs the consignee that GE HealthCare will replace the affected products with CRITIKON SOFT-CUF Blood Pressure Cuffs and a GE HealthCare representative will contact the consignee to arrange for this correction. An Acknowledgement Form was enclosed for completion and return.
Initiated
2024-04-09
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, LA, MD, MO, MS, NC, NJ, NM, OH, PA, TN, and TX.
Recall number
Z-2027-2024
Recall status
Open, Classified
CDRH record ID
207673
FDA classified
2024-06-06
Recall event number
94597

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.