Z-2028-2024: Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Ma

Recall record Z-2028-2024 for Remote Diagnostic Technologies Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2028-2024
Company in FDA record
Remote Diagnostic Technologies Ltd.
FDA classification date
Jun 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

Source facts

Action
On 5/9/24, Philips distributed correction notices to customers and distributors asking them to do the following: 1) The User/Operator Manual was updated can be downloaded by searching "Tempus LS-Manual User-Operator Manual" at Philips.com/IFU 2) This notice needs to be shared with those who need to be aware within your organization and any organization where the potentially affected devices have been transferred. 3) Complete and return the response form via email to ECR.Recall.Response@Philips.com. The following URL can also be used to complete/return the response form: https://philips.efrnfeedback.com/se/705E3ED80E91C689
Initiated
2024-05-09
Root cause
Labeling design
Distribution
US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT.
Recall number
Z-2028-2024
Recall status
Open, Classified
CDRH record ID
207740
FDA classified
2024-06-06
Recall event number
94618

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗