Z-2028-2025: REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500

Recall record Z-2028-2025 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2028-2025
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Jun 25, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flagged results being auto validated and uploaded to LIS.

Source facts

Action
On 04/07/2025, the firm sent via email, or postal mail an "Urgent Medical Device Recall" Letter to customers informing them that Beckman Coulter has identified an issue with REMISOL Advance when receiving a sample order from the Laboratory Information System (LIS) if the patient's Date of Birth (DoB) is not entered or is unknown. Customers are instructed to: . Verify that their system is configured with the default validation rules and ranges. REMISOL Advance will utilize these default validation rules in situations where the patient's age or date of birth is unknown (please see the chapter Reference and Validation Ranges 1-233 in the IFU) OR . Ensure that the DoB field is not left empty and confirm with their LIS vendor that they do not use 01/01/1900 as a default date of birth. If they do, request that this default date be changed to 02-Jan-1900 (02/01/1900) to allow REMISOL Advance to correctly estimate the patient's age. Beckman Coulter will: . The IFU will be revised to specify that the DoB is required for generating and validating results. . Beckman Coulter will provide software updates upon customer request. For HL7 Host Driver Users: . Beckman Coulter has implemented the correction in software versions = 2.3.09. For ASTM Host Driver Users: . Beckman Coulter will implement the correction in software version 2.4. For questions, contact Customer Support Center via http://www.beckmancoulter.com
Initiated
2025-04-08
Root cause
Software design
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Australia, Austria, Bangladesh, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Czechia, Denmark, Ethiopia, France, French Polynesia, Germany, Greece, Guam, Hong Kong, India, Ireland, Italy, Japan, Korea, Republic of, Kuwait, Kyrgyzstan, Lebanon, Malaysia, Mexico, Morocco, Netherlands, New Zealand, North Macedonia, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, Turkey, Uganda, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, and Zimbabwe
Recall number
Z-2028-2025
Recall status
Open, Classified
CDRH record ID
213399
FDA classified
2025-06-25
Recall event number
96685

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗