Skip to content

Public FDA record

Z-2029-2024: OES 4000 Hysteroscope, Model/Catalog Number: A4674A

Recall record Z-2029-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2029-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Jun 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

Source facts

Action
On May 9, 2024, URGENT: MEDICAL DEVICE REMOVAL letters were sent to the consignee. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Therefore, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Examine your inventory and identify the above listed device with serial number. 3. Please contact Customer Service at at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus, and we will replace this device at no charge. 4. Olympus requests that you acknowledge receipt of this letter through the a. Go to https://olympusamerica.com/recall b. Enter the recall number "XXXX . c. Complete the form as instructed and include your account ID number: XXXXXXXX 5. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you report any complaints, including foggy image, to the Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@olympus.com
Initiated
2024-04-18
Root cause
Process change control
Distribution
US Nationwide distribution in the state of NY.
Recall number
Z-2029-2024
Recall status
Open, Classified
CDRH record ID
207909
FDA classified
2024-06-06
Recall event number
94638

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗