Skip to content

Public FDA record

Z-2030-2009: Medtronic Physio-Control LIFEPAK CR Plus Defibrillator/Monitor. Automated external defibrillator deploye

Recall record Z-2030-2009 for Physio Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2030-2009
Company in FDA record
Physio Control, Inc.
FDA classification date
Sep 14, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The K1 relay on the Analog printed circuit board assembly may short due to moisture ingress in a highly humid environment. K1 shorting may affect the ECG amplitude causing the Shock Advisory System(SAS) to not analyze the ECG rhythm correctly thus preventing the device from delivering therapy.

Source facts

Action
Physio-Control, Inc. notified Customers by phone beginning August 18, 2009. In addition, the recalling firm Issued an "Urgent Medical Device Recall" letter dated August 2009 informing Customers of the affected product and informing them that a Physio-Control representative will call and arrange for immediate device replacement. For further information, contact Physio-Control Customer Care at 1-800-442-1142.
Initiated
2009-07-31
Root cause
Nonconforming Material/Component
Terminated
2013-03-15
Distribution
Worldwide Distribution -- US (DC and MN) and Japan.
Recall number
Z-2030-2009
Recall status
Terminated
CDRH record ID
84494
FDA classified
2009-09-14
Recall event number
53035

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗