Open original FDA record ↗Recall record
- FDA identifier
- Z-2030-2009
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Sep 14, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The K1 relay on the Analog printed circuit board assembly may short due to moisture ingress in a highly humid environment. K1 shorting may affect the ECG amplitude causing the Shock Advisory System(SAS) to not analyze the ECG rhythm correctly thus preventing the device from delivering therapy.
Source facts
- Action
- Physio-Control, Inc. notified Customers by phone beginning August 18, 2009. In addition, the recalling firm Issued an "Urgent Medical Device Recall" letter dated August 2009 informing Customers of the affected product and informing them that a Physio-Control representative will call and arrange for immediate device replacement. For further information, contact Physio-Control Customer Care at 1-800-442-1142.
- Initiated
- 2009-07-31
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-03-15
- Distribution
- Worldwide Distribution -- US (DC and MN) and Japan.
- Recall number
- Z-2030-2009
- Recall status
- Terminated
- CDRH record ID
- 84494
- FDA classified
- 2009-09-14
- Recall event number
- 53035