Open original FDA record ↗Recall record
- FDA identifier
- Z-2030-2012
- Company in FDA record
- Sigma
- FDA classification date
- Jul 18, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class III
Reason stated in the source
Sigma is a safety alert for the Master Drug Library (MDL) Editor software associated with the SIGMA Spectrum Infusion Pump System because it may result in a facility drug library that does not represent common clinical practices.
Source facts
- Action
- Sigma sent an Urgent: Customer Safety Alert dated April 23, 2012 via UPS delivery to all consignees requesting both acknowledgement of notification and confirmation that recommended actions have been taken. Customers are asked to follow the instructions contained in the recall letter. Customers with questions may contact the firm at 800-356-3454 ext 300.
- Initiated
- 2012-04-23
- Root cause
- Software design
- Terminated
- 2018-07-05
- Distribution
- Worldwide Distribution - US (nationwide) including Puerto Rico, and Canada.
- Recall number
- Z-2030-2012
- Recall status
- Terminated
- CDRH record ID
- 108723
- FDA classified
- 2012-07-18
- Recall event number
- 61691