Open original FDA record ↗Recall record
- FDA identifier
- Z-2031-2019
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Jul 22, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Source facts
- Action
- Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers. In the event the issue occurs, the system cannot be operated normally as X-ray will not possible. Planned procedures may have to be terminated and performed on an alternative X-ray system. Siemens advises that an alternative system is available to continue treatment in case of occurrence. Siemens is implementing a corrective action to all potentially affected users of the Artis Zee and AXIOM Artis systems with generator A100 via Update Instruction AX038/18/S. The Siemens Service organization will modify the affected supply circuit and revise the wiring. This modification will bring the voltage back to the center of the tolerance range. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated and any recurrence of this potential fault is prevented.
- Initiated
- 2019-06-12
- Root cause
- Component design/selection
- Terminated
- 2020-09-18
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2031-2019
- Recall status
- Terminated
- CDRH record ID
- 174028
- FDA record created
- 2019-07-22
- Recall event number
- 83203