Z-2032-2014: Zimmer EDI PERIARTICULAR PLATING SYSTEM CANCELLOUS BONE SCREW, FULLY THREADED, HEX HEAD 4MM DIAMETER,

Recall record Z-2032-2014 for Zimmer Manufacturing B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2032-2014
Company in FDA record
Zimmer Manufacturing B.V.
FDA classification date
Jul 10, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two lots of screws were commingled. Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number 62628759; and screws with etching and machining for lot number 62628759 were packaged in bags labeled for lot number 62629432. The affected product was distributed between 3/6/14 and 4/21/14.

Source facts

Action
Zimmer sent an Urgent Medical Device Recall Lot specific letter dated June 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: 1. Please locate, quarantine, and return the affected product (part number 00-4840-020-00 lot number 62629432 and part number 00-2347-022-14 lot number 62628759) from all inventories. 2. Please note that any hospitals that received direct shipments of this product from Zimmer will be notified via Fed Ex. 3. Please review the list of hospital and surgeons to be provided to your territory via email. 4. Provide the name, address, and phone number of any additional hospitals or surgeons that may have received or used the affected product, or confirm that the list provided is complete. 5. Complete and return Attachment 1. 6. Ensure that a copy of the Risk Managers and Surgeons letter is left with the facilities from which product is removed by providing them with a copy of the letter, if they are not identified as a facility receiving a letter directly from Zimmer. 7. If after reviewing this notification you have questions or concerns, please contact Zimmer at 1-877-946-2761.
Initiated
2014-06-12
Root cause
Process design
Terminated
2015-06-24
Distribution
Worldwide Distribution - USA (nationwide) and the countries of India, Japan, China, Australia and Canada.
Recall number
Z-2032-2014
Recall status
Terminated
CDRH record ID
128171
FDA classified
2014-07-10
Recall event number
68547

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗