The 8-inch model of the Jewel Precision Reusable Rigid Sterilization Container System (i.e., model number: JP-24-8) was not cleared by the FDA. The letter to file justifying the change in size was insufficient.
Source facts
Action
Jewel Precision sent a "MEDICAL DEVICE RECALL" letter dated 5/23/2025 to the single consignee on 05/23/2025 via email. The consignee was instructed to; review all stock and inventory, return the product to Jewel Precision for destruction, complete the Acknowledgement and Receipt form on pages 3-4 of the letter indicating the actions taken and approval of completion, in the event that the product has been scrapped, please provide evidence of scrapping it and certification that the product has been destroyed by the customer. As per the customer's purchase order, the piece was to be used for testing purposes only.
For products that have been shipped to your customers, identify and notify your customers of the recall, requesting return; obtain their acknowledgement and approval that the part has been shipped back.
Action by company: discontinued the product.
For questions email: jsitler@jewelprecision.com or call 973-857-5545 x19.
Initiated
2025-05-23
Root cause
No Marketing Application
Distribution
US distribution to Tennessee.
Recall number
Z-2033-2025
Recall status
Open, Classified
CDRH record ID
214109
FDA classified
2025-06-26
Recall event number
96958
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.