Customer reported problem while performing incoming QC of newly received lot of test cards. Liquid quality control results were found to be out of range. Further investigation by Epocal determined that a portion of the test card lot (approximately 3000 cards) could potentially report low glucose results.
Source facts
Action
Epocal Inc. sent a Voluntary Product Recall letter dated June 27, 2011 to the affected consignee. The letter identified the affected products, problem and actions to be taken. The letter ask that all products affected by this recall be sent back to Epocal and a replacement may be obtained. For questions call 1-613-738-6912.
Initiated
2011-06-28
Root cause
Device Design
Terminated
2013-12-18
Distribution
US nationwide Distribution in the state of CA.
Recall number
Z-2034-2012
Recall status
Terminated
CDRH record ID
107399
FDA classified
2012-07-19
Recall event number
61147
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.