Open original FDA record ↗Recall record
- FDA identifier
- Z-2034-2025
- Company in FDA record
- Philips North America Llc
- FDA classification date
- Jun 26, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.
Source facts
- Action
- An "URGENT Medical Device Correction" notice dated 5/21/25 was mailed to consignees. The notification instructs consignees on how to identify if their devices have affected software downloaded. Consignees may continue to use affected systems in accordance with the intended use and recommended actions provided in the recall notification in Appendix A. Appendix A summarizes the ten software issues and the actions recommended by the manufacturer that the consignee follow. Consignees are to complete and return the provided response form, forward the recall notification to users, and retain a copy with all affected systems until a solution is installed. Philips will contact consignees to schedule time for a Field Service Engineer to install the solution and resolve the issue (Reference FCO72800828/FCO 72800829). Consignees with any questions can contact Customer Care Solutions at 1-800-722-9377.
- Initiated
- 2025-05-29
- Root cause
- Under Investigation by firm
- Distribution
- Domestic: AZ, CA, CO, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV; Foreign: Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Lithuania, Macao, Malaysia, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.
- Recall number
- Z-2034-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214212
- FDA classified
- 2025-06-26
- Recall event number
- 97010