Z-2034-2025: Spectral CT. Computed tomography X-ray system.

Recall record Z-2034-2025 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2034-2025
Company in FDA record
Philips North America Llc
FDA classification date
Jun 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.

Source facts

Action
An "URGENT Medical Device Correction" notice dated 5/21/25 was mailed to consignees. The notification instructs consignees on how to identify if their devices have affected software downloaded. Consignees may continue to use affected systems in accordance with the intended use and recommended actions provided in the recall notification in Appendix A. Appendix A summarizes the ten software issues and the actions recommended by the manufacturer that the consignee follow. Consignees are to complete and return the provided response form, forward the recall notification to users, and retain a copy with all affected systems until a solution is installed. Philips will contact consignees to schedule time for a Field Service Engineer to install the solution and resolve the issue (Reference FCO72800828/FCO 72800829). Consignees with any questions can contact Customer Care Solutions at 1-800-722-9377.
Initiated
2025-05-29
Root cause
Under Investigation by firm
Distribution
Domestic: AZ, CA, CO, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV; Foreign: Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Lithuania, Macao, Malaysia, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.
Recall number
Z-2034-2025
Recall status
Open, Classified
CDRH record ID
214212
FDA classified
2025-06-26
Recall event number
97010

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗