Open original FDA record ↗Recall record
- FDA identifier
- Z-2035-2008
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Sep 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Based on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a patient, there is a risk of infection. If handle discharge button is damaged during sterilization, defibrillation therapy may be prevented.
Source facts
- Action
- Starting 2/5/08, Physio-Control sent letters to customers which notifies them not to use Flash Gravity Steam, 12/88 Ethylene Oxide, STERRAD 100 machine, or Gravity Steam to sterilize. The firm recommends the use of 100% EO, STERRAD 100S machine, or Prevacuum steam for sterilization. The letter also recommends that after sterilization, the handles and paddles should be examined carefully for damage. If damaged, the product should be removed from use. The firm will follow up with a revised Sterilization Guideline document and requests posting the letter in the sterilization department in the meantime. Contact technical support at 800-442-1142 if you have questions. On 7/23/08 the firm issued new sterilization guidelines that supersede all others.
- Initiated
- 2008-01-29
- Root cause
- Labeling design
- Terminated
- 2011-05-04
- Distribution
- Distributed to hospitals and medical centers. Nationwide, DOD, and international: ARGENTINA, BAHRAIN, BANGLADESH, BERMUDA, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, INDIA, IRAN, ISRAEL, JAPAN, KOREA, LEBANON, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, QATAR, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SRI LANKA, SYRIA, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIRGIN ISLANDS - US, YEMEN, ZIMBABWE
- Recall number
- Z-2035-2008
- Recall status
- Terminated
- CDRH record ID
- 68214
- FDA classified
- 2008-09-16
- Recall event number
- 46749