The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.
Source facts
Action
On May 3, 2018, the firm issued an Urgent Amended Medical Device Correction notice to affected customers via certified mail. Customers were reminded that they were previously notified in April 2016 regarding the same issue. At that time, the firm committed to providing a replacement for all affected MRx units. The May 2018 notification was sent only to remaining customers who had not yet received a replacement device. Those customers were informed that due to the product being discontinued, their affected units would receive a replacement therapy board instead of a replacement unit.
Customers were informed that they could continue to use their MRx prior to repair. But, if they observe a red X in the Ready for Use indicator, to follow the troubleshooting instructions in the MRx IFU. They were also advised that they would be contacted by Philips to schedule the repair of the device. If the customer has already returned their unit but has not yet received a replacement, they were advised that Philips will be replacing the therapy board, not the entire device.
If you need further information or support concerning this notification,
please contact your local Philips representative or call the firm at 1-800-722-
9377.
Initiated
2018-05-03
Root cause
Employee error
Terminated
2019-10-17
Distribution
Worldwide Distribution to accounts in OH and NC. Foreign distribution to Lebanon.
Recall number
Z-2035-2018
Recall status
Terminated
CDRH record ID
164128
FDA record created
2018-05-30
Recall event number
79982
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.