When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not operational for clinical use. This problem occurs randomly.
Source facts
Action
Philips Healthcare is informing the consignee of this issue through a Field Safety Notice (FSN) 88200487-488. The US consignees will receive this notice via certified mail and Philips Key Market representatives will distribute the FSN internationally. Additionally, a Field Service Engineer will contact each site to schedule a time to install the software update.
Initiated
2014-06-18
Root cause
Software design
Terminated
2015-09-16
Distribution
Distributed in the states of CA, CO, IL, OK, OR and WI.
Recall number
Z-2037-2014
Recall status
Terminated